How to set PEEP when using emergency ventilator JIXI-H-100A

Release:jiuxin-medical Browse:668

JIXI-H-100A is a trolley-type emergency ventilator manufactured by us.

Now we will introduce how to set PEEP when using the model.

 

The range of PEEP is 0, 5-20hPa.

During exhalation period, the ventilator opens the exhalation channel according to the set PEEP data.

 

Note:

l The lowest PEEP set data should be higher than the present trigger set data, and the highest PEEP set data should be 5hPa lower than the present peak airway pressure.

l When adjust PEEP, the tidal volume and oxygen concentration will be changed, and please modify in time.  