How to set PEEP when using emergency ventilator JIXI-H-100A

Release:jiuxin-medical Browse:247

JIXI-H-100A is a trolley-type emergency ventilator manufactured by us.

Now we will introduce how to set PEEP when using the model.

 

The range of PEEP is 0, 5-20hPa.

During exhalation period, the ventilator opens the exhalation channel according to the set PEEP data.

 

Note:

l The lowest PEEP set data should be higher than the present trigger set data, and the highest PEEP set data should be 5hPa lower than the present peak airway pressure.

When adjust PEEP, the tidal volume and oxygen concentration will be changed, and please modify in time.  


ADD:No.168, Fenghuang West Road, Taizhou, Jiangsu Province, China.

TEL:+86 523-86880973-ext

FAX:+86 523-86196046

QQ :2256378460

MAIL:info@jiuxin-medical.com

jiuxin-medical@163.com

Justsends Medical Technology(Taizhou) Co.,Ltd.     苏ICP备11030010号-1