Knowledge of oxygen cylinder inflation for emergency ventilator

jiuxin-medical 16times

JIUJIUXIN emergency transport ventilator is equipped with 2L aluminum alloy oxygen cylinder as standard configuration.

Advantages of aluminum alloy oxygen cylinder for transport ventilator: light weight, high pressure resistance, long-term use of anhydrous rust and impurities.

The working pressure range of aluminum alloy cylinder is 15MPa (independent of cylinder size).

Inflation procedure of oxygen cylinder for transport/emergency/portable ventilator:

Close the valves of large and small oxygen cylinders - connect the oxygen bridge to the small one - connect the large one to the other end of the oxygen bridge - open the valve of the small oxygen cylinder - slowly open the valve of the large oxygen cylinder - finish the inflation - close the large oxygen cylinder first - close the small oxygen cylinder again.

Attention:

1. It is normal that the oxygen cylinder is heated or accompanied with air flow sound during inflation;

2. When the air flow sound stops, it means that the pressure on both sides is basically balanced and the inflation is completed;

3. After inflation, turn off the switches of the big and small oxygen cylinders respectively, and then remove the inflation oxygen bridge;

4. After inflation, connect the pressure reducing valve to check the pressure in the oxygen cylinder;

5. No fireworks when inflating.


ADD:No.168, Fenghuang West Road, Taizhou, Jiangsu Province, China.

TEL:+86 523-86880973-ext

FAX:+86 523-86196046

QQ :2256378460

MAIL:info@jiuxin-medical.com

jiuxin-medical@163.com